NVPC
Online
NVPC: Make= Beckman; Model= Met one 6015P; Capacity= 1CFM
SOFTWARE= EnVigil
Critical
alarm: Critical
alarms are those alarms when generated indicate excursion which may impact the
product quality. Alarm generated in location where the NVPC probe is in close
proximity to the product/ sterile container closure/ product contact path.
Location for critical alarms includes
grade A location - Filling station, Bunging station, Bunging Hopper station,
Portable cart, IOS station, AGV station, Near lyophilizer door open, between
lyophilizers (near Lyo) and Lyo LAF (rigid barrier transfer conveyor).
Critical alarms include the following-
1.
Particle
count excursion beyond the acceptance criteria at a particular location.
2.
Sensor
Failure, Counter Flow Failure, ERROR or SKIPPED message.
Ensure and verify the status of counter
flow to meet the below set limit-
Target Counter flow – 1.0 CFM.
Minimum counter flow: 0.9 CFM.
Maximum counter flow: 1.1 CFM.
“0” states counter ‘OFF’ condition.
“1” states counter ‘ON’ condition.
GRADE- A (On Line)
0.5 Micron= 100 (Cubic feet/min) and 3520 (Cubic meter/35.31 minutes)
5 Micron= 0 (Cubic feet/min) and 20 (Cubic meter/35.31 minutes)
GRADE- A (On Line)
0.5 Micron= 100 (Cubic feet/min) and 3520 (Cubic meter/35.31 minutes)
5 Micron= 0 (Cubic feet/min) and 20 (Cubic meter/35.31 minutes)
GRADE- B
0.5 Micron= 352000 (Cubic meter)
5 Micron= 2900 (Cubic meter)
GRADE- C
0.5 Micron= 3520000 (Cubic meter)
5 Micron= 29000 (Cubic meter)