Dry heat sterilization


PDA Technical report No. : 3 (Revised 2013)
Dry heat process used for sterilization and depyrogenation.
Bacterial Endotoxin Test (BET) - Assay for measuring active endotoxins by combining a liquid test sample with LAL (Limulus Amoebocyte Lysate) reagent and measuring the resulting proportional reaction via visual, turbidimetric, chromogenic, or other validated means of detection.
BI Challenge system - A test system containing viable microorganisms of pure and specified strain providing a defined resistance to a specified sterilization process.
Depyrogenation - Destruction and /or removal of bacterial endotoxins. It demonstrates at least 99.9 % or 3 Log endotoxins reduction.
D- Value- Time in minutes required for a one-logarithm or 90% reduction of microorganisms used as biological indicator under specified lethal conditions.
Dwell Time- Period that items are subjected to a given processing conditions.
Endotoxin Indicator (EI) for depyrogenation - Vial of endotoxin (or a carrier spiked with endotoxin) designed for use in depyrogenation studies. Endotoxin on a carrier is added at a concentration sufficient to allow recovery of a minimum of 1000 USP endotoxin units/ carrier. The endotoxin indicator would allow for accurate indication of at least a 3 Log reduction in USP endotoxin units during depyrogenation process challenges.


Tunnel Biological indicator ======Fix the thermocouples along with Biological indicator (Bacillus Atrophaeus) on the cooling chamber internal surface/belt by opening

Comments

Popular posts from this blog

Change Management System

US Market (Supplement): Post approval drug manufacturing supplement for NDAs/ ANDA submitted to FDA for Major or Moderate manufacturing change. Major change - If a manufacturing change is considered to be major, an applicant must submit and receive FDA approval of a prior-approval supplement (PAS) before the drug product made with the change is distributed. Moderate change- If a manufacturing change is considered to be moderate, an applicant must submit a supplement at least 30 days before the drug product is distributed (a CBE-30 supplement) or, in some cases, submit a supplement at the time of distribution (a CBE-0 supplement).  “CBE” means “changes-being-effected”. Post manufacturing supplement - 1. CBE 0 (Change being effected before 00 days): 2. CBE 30 (Change being effected before 30 days): 3. PAS (Prior approval supplement): EUROPE Market (Variation): A variation to the terms of a marketing authorization is an amendment to the contents of the docume...

NVPC

NVPC Online NVPC: Make= Beckman; Model= Met one 6015P; Capacity= 1CFM SOFTWARE= EnVigil Critical alarm: Critical alarms are those alarms when generated indicate excursion which may impact the product quality. Alarm generated in location where the NVPC probe is in close proximity to the product/ sterile container closure/ product contact path. Location for critical alarms includes grade A location - Filling station, Bunging station, Bunging Hopper station, Portable cart, IOS station, AGV station, Near lyophilizer door open, between lyophilizers (near Lyo) and Lyo LAF (rigid barrier transfer conveyor). Critical alarms include the following- 1.     Particle count excursion beyond the acceptance criteria at a particular location. 2.     Sensor Failure, Counter Flow Failure, ERROR or SKIPPED message. Ensure and verify the status of counter flow to meet the below set limit- Target Counter flow – 1.0 CFM. Minimum counter flow: ...

Air velocity measurement and correlation to smoke study

Air velocity measurement As part of risk assessment this evaluation should also consider the selection of locations for the critical control points for monitoring air velocity as well as viable and non-viable particles. 1. Problem Statement- Where should airflow velocity measurements be taken with respect to a filling line or other aseptic processing areas? Recommendation Airflow velocity measurements should be taken at locations where meaningful and reproducible results can be obtained. This is typically at a distance of 6 inches from the filter face . Rationale for Recommendation The primary reason for airflow velocity measurements in unidirectional airflow areas (e.g., areas where product, product contact packaging components, and product contact surfaces are exposed) is to assure there is adequate airflow to protect the materials from external airborne contamination . Accurate measurements can be taken and changes over time detected when airflow velocities are e...